Abstract
A neutron-detection element using circular cathode bumps was developed for two-dimensional neutron measurement, and an irradiation experiment was performed using a Cf-252 neutron source. The element has triangularly arranged small-sized circular cathode bumps, and the bumps are linked together in the x- and y-directions for the detection of incident neutrons. Preliminary irradiation experiments for the developed element were performed using a neutron-detection system consisting of a pressure vessel, amplifier-shaper-discriminator boards, optical signal transmission devices, position encoders, and a data acquisition device. The element was arranged in the pressure vessel with a fill gas of composition He/(15%)CF4 at 0.7 MPa. The detector system exhibited a two-dimensional uniformity with a response of 10.1%. The average intrinsic spatial resolution was 1.89 mm full width at half maximum in the sensitive region, calculated by taking into account the track lengths of the secondary particles.
Published Version
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.