Abstract
Coronary plaque morphology, including plaque size and fibrous cap thickness, is thought to contribute to the risk of plaque rupture and future cardiac events. Dual-frequency intravascular ultrasound has been proposed as a possible technique to visualize both large-scale features and superficial detail of coronary plaque; however, it has not been found to be feasible within the constraints of a clinically functional intravascular ultrasound catheter. In this study, we describe the design and fabrication of a dual-frequency catheter using a bidirectional transducer stack with center frequencies of approximately 30 and 80 MHz. We describe how the high-frequency transducer achieves significantly improved axial and lateral resolution (16 and 120 µm, respectively, vs. 50 and 220 µm) at the expense of penetration depth. Finally, imaging of ex vivo human coronary artery segments reveals that the catheter can provide complementary images of the deeper arterial wall and superficial plaque features.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.