Abstract

AbstractMexiletine (MEX) is an antiarrhythmic drug used to treat acute and chronic ventricular arrhythmias. A simple isocratic, rapid, specific, stability‐indicating, and sensitive reverse‐phase ultra‐high‐performance liquid chromatography (UHPLC) method was developed and successfully validated to quantify the assay and impurities of MEX hydrochloride. The new UHPLC method showed the optimum separation of MEX and its seven impurities in 20 min by using sodium acetate buffer (0.14 mM) with 0.3% of glacial acetic acid (pH 4.8) and methanol in the ratio of 40:60 (v/v) as mobile phase. The method used an Acquity HSS T3 (100 × 3.0 mm, 1.8 µm) chromatographic column at a 0.2 mL/min flow rate. The injection volume is 2 µL, and 254 nm is the detection wavelength. Note that, 25°C is the column oven temperature. The method is linear over the concentration range of the quantification limit (0.03%) to 150% (0.3%) of the specification level (0.2%) for the impurities and 50%–150% for the assay method. The quality control lab of bulk drugs and finished dosage form successfully applied the new UHPLC method for quantification of assay and impurities of MEX hydrochloride.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call