Abstract
In the present paper, sensitive, rapid, and different analytical methodology was developed for the determination of anticancer drug epirubicin (EPR). The mixture of epirubicin and moxifloxacin as internal standard was separated on a reversed phase Waters Spherisorb ODS1 column (250mm × 4.6mm × 5mm) using acetonitrile/water (30:70 v/v) mixture containing 15 mM phosphoric acid as mobile phase at 0.6 mL min−1 flow rate and 30 C. Also degradation studies were conducted as stress conditions of UV light, acidic hydrolysis, alkaline hydrolysis, oxidation, and heat in oven (100°C), to evaluate the ability of the proposed method for the separation of EPR from its degradation products. The validated method suggests routine analysis of EPR in differently equipped laboratories.
Published Version
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have