Abstract

A simple, precise, and accurate hydrophilic interaction liquid chromatographic (HILIC) method has been developed for the determination of five aromatic amines in environmental water samples. Chromatography was carried out on a bare silica column, using a mixture of acetonitrile and a buffer of NaH(2)PO(4)-H(3)PO(4) (pH 1.5, containing 10mM NaH(2)PO(4)) (85:15, v/v) as a mobile phase at a flow rate of 1 mL min(-1). Aromatic amines were detected by UV absorbance at 254 nm. The linear range of amines was good (r(2)>0.998) and limit of detection (LOD) within 0.02-0.2 mg L(-1) (S/N=3). The retention mechanism for the analytes under the optimum conditions was determined to be a combination of adsorption, partition and ionic interactions. The proposed method was applied to the environmental water samples. Aromatic amines were isolated from aqueous samples using solid-phase extraction (SPE) with Oasis HLB cartridges. Recoveries of greater than 75% with precision (RSD) less than 12% were obtained at amine concentrations of 5-50 microg L(-1) from 100mL river water and influents from a wastewater treatment plant (WWTP). The present HILIC technique proved to be a viable method for the analysis of aromatic amines in the environmental water samples.