Abstract
This investigation is aimed to develop and validate a high-performance thin layer chromatography method for quantitative determination of serum zonisamide levels. Chromatographic separation was carried out on a silica gel 60F254 plate using a mixture of ethyl acetate:methanol:toluene (4:1:5) as the mobile phase. Densitometric detection was carried out at 254 nm. The method was validated for linearity, precision, selectivity, limit of detection, limit of quantification and accuracy. Linear calibration curves in the range of 5 to 80 µg/band gave a correlation coefficient of 0.991. The intra-day (n=6) and inter-day (n=18) precision, expressed as the relative standard deviation were in the range of 2.83 to 3.27 % and from 2.09 to 4.39 %. The limits of detection and quantification were found to be 1.07 and 3.15 µg/band, respectively. Accuracy was calculated as percent recovery and was found to be 97.52 and 115.23 %. Theophylline was used as an internal standard, which gave a well separated peak at Rf 0.36 without interfering with zonisamide. The method was found to be specific with no matrix interference. Thus, the method developed for the estimation of serum zonisamide level is simple, cost-effective and reliable for therapeutic drug monitoring.
Published Version (Free)
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.