Abstract
An isocratic high performance liquid chromatographic method has been developed and validated for the estimation of benzoyl peroxide in flours.The benzoyl peroxide was extracted with acetonitrile and analysed by Waters HPLC instrument on a C18 column using water-acetonitrile mobile phase with a flow rate of 1.5 ml/min and UV detection at 235 nm.The linearity was found in the range of 5-50 mg/l.The method is simple,reliable, specific,accurate and precise. KEY WORDS: Benzoyl peroxide, Acetonitrile, HPLC, Method Development,Method Validation,.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: EPRA International Journal of Multidisciplinary Research (IJMR)
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.