Abstract

Comprehensive two-dimensional liquid chromatography (LC × LC) is a powerful tool for the separation of complex biological samples. This technique offers the advantage of simplified automation and greater reproducibility in a shorter analysis time than off-line two-dimensional separation systems. In the present study, an LC × LC system is developed enabling simultaneous UV and MS detection, and which can be easily converted to a conventional reversed-phase LC-UV/MS system. In LC × LC, a 60-min reversed-phase LC separation with a linear solvent gradient in the first dimension is coupled to a second-dimension separation on a mixed-mode cation-exchange/reversed-phase column with a modulation time of 60 s. The isocratic separation in the second-dimension column is optimized by the use of a multi-step gradient where the organic and the ionic modifier are varied independently. Intraday ( n = 3) and interday ( n = 4) variability of the retention times were evaluated with the complete system and found to be 0.5% and 0.7%, respectively. Good linearity was observed in calibration curves for three different compounds varying in polarity.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.