Abstract

A selective, suitable and reproducible HPLC-UV method was developed and validated for the determination and pharmacokinetic investigation of valsartan in human plasma. Valsartan and internal standard (IS), irbesartan, were extracted from plasma samples using liquid–liquid extraction with ethyl acetate + n-hexane (80:20). Chromatographic separation was performed on a Lichrosphere C18 RP Select B column (250 × 4.0 mm i.d; 5 µm particle size) using 20-mM potassium dihydrogen orthophosphate buffer (pH adjusted to 2.7 ± 0.05 using 50% orthophosphoric acid):acetonitrile (60:40 v/v) as the mobile phase at flow rate of 1.2 ml/min. The effluent was monitored using ultraviolet (UV) detection set at 225 nm, having column oven temperature 40°C and sample cooler temperature 8°C ± 0.2°C. The calibration curve was linear over the range of 217.7–6118.4 ng/mL. The intra- and inter-day precisions were within 8.6% and 7.4%, respectively, whereas accuracies were 101.0–107.9% and 103.4–109.5%, respectively. The validated meth...

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.