A novel x-ray Bragg magnifier of variable magnification is proposed. We describe the basic properties of the magnifier based on the dynamical theory of x-ray diffraction, and experimentally investigated its performance at the vertical wiggler beamline BL-14B of the Photon Factory. Using a Si(220) asymmetric crystal as the Bragg magnifier, we successfully controlled the magnification factor, M, between 0.1 and 10.0 at a wavelength of 0.112nm. The observed spatial resolution with an x-ray CCD camera, the pixel size of which was 23μm×23μm, was 10μm at M=10, 50μm at M=1 and 250μm at M=0.17. As a demonstration, we successfully used the x-ray magnifier for computed tomography for the non-destructive observation of a glass tube at several magnification factors.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.