We report on the development of a SQUID read-out for a fast cryogenic detector with superconducting phase transition thermometer. To avoid the usual transformer matching and modulation–demodulation technique we are using a second SQUID to amplify the signal of a first stage input SQUID. A bandwidth of 2 MHz and a slew rate of 2×10 5 Φ 0 s −1 were achieved. The system is developed for the detection of large protein molecules in a time of flight mass spectrometer.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.