Abstract

A high pressure vessel equipped with continuous magnetic stirring has been interfaced directly online with supercritical fluid chromatography (SFC). The interface consists of an oven housing a recirculation pump that maintains a constant flow of supercritical fluid solution from the vessel through a sample injection valve for online SFC analysis. The method involves increasing the vessel pressure in controlled steps to produce a series of progressively more saturated supercritical fluid carbon dioxide solutions for online SFC solubility measurements. The integrated static solubility system provides a convenient means to ensure equilibrium is established prior to performing online SFC solubility measurements and also facilitates a self-validation accuracy check procedure. For nifedipine, solubility data was obtained at 333.15 and 353.15 K through the pressure range from 12.5 to 27.5 MPa using 2.5 MPa steps. For quinine, solubility data was obtained at 323.15 and 343.15 K through the pressure range from 12....

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.