Abstract

To quantify the resolution limits of scanning microwave impedance microscopy (sMIM), we created scanning tunneling microscope (STM)-patterned donor nanostructures in silicon composed of 10nm lines of highly conductive silicon buried under a protective top cap of silicon, and imaged them with sMIM. This dopant pattern is an ideal test of the resolution and sensitivity of the sMIM technique, as it is made with nm-resolution and offers minimal complications from topography convolution. It has been determined that typical sMIM tips can resolve lines down to ∼80nm spacing, while resolution is independent of tip geometry as extreme tip wear does not change the resolving power, contrary to traditional scanning capacitance microscopy (SCM). Going forward, sMIM is an ideal technique for qualifying buried patterned devices, potentially allowing for quantitative post-fabrication characterization of donor structures, which may be an important tool for the study of atomic-scale transistors and state of the art quantum computation schemes.