Abstract

Determining fluoride ions in ammonia desulphurization slurry using a fluoride ion selective electrode (ISE) is investigated. The influence of pH was studied and the appropriate total ionic strength adjustment buffer and its dosage were optimized. The impact of Fe3+ concentration on the detection results was analyzed under preferable conditions, and the error analysis of the ISE method’s accuracy and precision for measuring fluoride ion concentration in the range of 0.5-2000 mg/L was conducted. The quantitative recovery of F− in ammonium sulfate slurry was assessed. The results showed that when pH ranged from 5.5∼6 and the Fe3+ concentration was less than 750 mg/L, the accuracy and precision test results with quantitative recovery rates of 92.0%-104.2% were obtained.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.