Abstract

An efficient method for the determination of nanogram levels of zinc in waters (potable, irrigation, waste and sea waters) is described. Acidified water samples were analyzed by an accurately optimized time/temperature program for electrothermal atomic absorption spectrometry. The samples were previously treated with a matrix modifier consisting of 0.1% Mg(NO 3) 2 in 0.5 M HNO 3 and injected through a graphite tube with L’Vov platform. Sample recoveries, repeteability, and analyses of NIST reference material demonstrated the reliability and accuracy of this technique. Zinc concentrations ranged from 2.0 to 73.0 μg l −1. No significant differences among the mean zinc levels determined in the four types of water considered from southeastern Spain were found. Zinc concentrations in waters from the zone were not affected by human and industrial activity.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.