Abstract
For the determination of vitamin B 6 vitamers (pyridoxal phosphate, pyridoxamine phosphate, pyridoxal, pyridoxine, pyridoxamine) and 4-pyridoxic acid in biological samples such as plasma, cerebrospinal fluid and rat brain regions, a sensitive micromethod using high-performance liquid chromatography (HPLC) with fluorescence detection in combination with post-column derivatization is described. Metaphosphoric acid tissue extracts with deoxypyridoxine as an internal standard were injected into the HPLC system with a binary gradient elution at a flow-rate of 1.2 ml/min. The excitation wavelength of the fluorescence detector was set at 328 nm and the emission wavelength at 393 nm with a 15-nm slit width for the photocell. This method allows the assay of vitamin B 6 vitamers within 30 min in one chromatographic run. The present method has been applied extensively for the measurement of vitamin B 6 vitamer levels in discrete brain regions of small animals, cells in culture and biopsy samples.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Journal of Chromatography B: Biomedical Sciences and Applications
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.