Abstract

The authors report on a method to determine the Rabi frequency and transit time distribution of atoms that are essential for proper operation of atomic beam frequency standards. Their method, which employs alternative regularized inverse on two Ramsey spectra measured at different microwave powers, can be used for the frequency standards with short Ramsey cavity as well as long ones. The authors demonstrate that uncertainty in Rabi frequency obtained from their method is 0.02%.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.