Abstract

AbstractThis work reports the trace determination of Tl(I) by square‐wave anodic stripping voltammetry (SWASV) on novel microsensors equipped with a bismuth‐film electrode (BiFE). The sensors were fabricated by a multistep microfabrication approach combining sputtering (to deposit the electrode materialm, bismuth‐ and the insulator SiO2, on the surface of a silicon wafer) and photolithography (to define the geometry of the sensor). The effect of the preconcentration time, the preconcentration potential and the SW stripping parameters were investigated. Using the selected conditions, the 3σ limit of detection was 0.6 µg L−1 of Tl(I) at a preconcentration time of 240 s and the percent relative standard deviation was 4.3 % at the 10 µg L−1 level (n=8). In order to eliminate the interference caused by Pb(II) and Cd(II), EDTA was added in the sample solution The method was successfully applied to the determination of Tl(I) in a certified lake water sample. These new sensors exhibit excellent mechanical stability and offer wide scope as mercury‐free disposable sensors for trace metal analysis.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.