Abstract

This work reports the simultaneous determination of Cd(II), Pb(II) and Zn(II) at the low μg l −1 concentration levels by square wave anodic stripping voltammetry (SWASV) on a bismuth-film electrode (BFE) plated in situ. The metal ions and bismuth were simultaneously deposited by reduction at −1.4 V on a rotating glassy carbon disk electrode. Then, the preconcentrated metals were oxidised by scanning the potential of the electrode from −1.4 to 0 V using a square-wave waveform. The stripping current arising from the oxidation of each metal was related to the concentration of each metal in the sample. The parameters for the simultaneous determination of the three metals were investigated with the view to apply this type of voltammetric sensor to real samples containing low concentrations of metals. Using the selected conditions, the limits of detection were 0.2 μg l −1 for Cd and for Pb and 0.7 μg l −1 for Zn at a preconcentration time of 10 min. Finally, BFE's were successfully applied to the determination of Pb and Zn in tapwater and human hair and the results were in satisfactory statistical agreement with atomic absorption spectroscopy (AAS).

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.