Abstract

Abstract An ion chromatography method has been proposed for the determination of three common inorganic anions (chloride, nitrate and sulfate) in anionic surfactants using a single pump system. The new system consists of an ion exclusion column, a concentrator column, and an anion exchange column connected in series via two 6-ports valves in a Dionex ICS-2000 ion chromatograph. The valves were switched several times for removing surfactants, concentrating and separating the three anions. The chromatographic conditions were optimized. Detection limits (S/N = 3) were in the range of 0.10–0.68 μg/L. The relative standard deviations (RSDs) of peak area were less than 4.6%. The recoveries were in the range of 84.1–112.6%.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.