Abstract
An ion chromatography (IC) system with switching recycling column was developed for the determination of bromate in drinking water with high chloride content. The system included a pump, two switching valves, guard column and analytical column, suppressor and a conductivity detector. In this method, the single eluent was used for both online collection and matrix elimination, and bromate was eluted from the concentrator column to the analytical column circularly. Under the optimized separation conditions, the method showed good linearity (r=0.9995) in the range of 1-100μg/L and repeatability (RSD≤4.80%, n=6). The limit of detection was 0.2μg/L (S/N=3) with 1000μL sample volume injections. The method was applied to analysis of drinking and tap waters, and satisfactory recoveries of spiked samples between 96.8% and 108.7% were obtained. The result showed that the recycling column-switching system can be useful for the determination of traces of bromate in high-chloride samples, which is required in production of the drinking water and quality control of the final product.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.