Total petroleum hydrocarbons (TPHs) in soil are determined by on-line dynamic supercritical fluid extraction (SFE) using infrared filter photometry detection. The filter photometer was constructed in the laboratory using a tungsten lamp, an optical notch filter that selects the C–H stretching vibration of the extracted organics, an optical chopper with demodulation electronics, and a PbSe detector. A modified high-pressure fiber optic flow cell was used to couple the SFE system to the photometer. Quantitation of TPHs was accomplished through the construction of calibration curves of integrated absorbance of C–H stretching (over time) versus concentration. Our studies show that the sensitivity of this system is affected by both the optical path length in the high-pressure cell and the SFE fluid flow-rate, and detection limits for TPHs are in the mid part-per-million range. The results of the application of this on-line SFE–IR instrument to the determination of TPHs in real-world samples show good agreement with those obtained from standard Soxhlet extraction–IR methods.