Abstract
A method is described for the flow injection determination of total iron as Fe(III) using a solid-phase reactor containing disodium-1,2-dihydroxybenzene-3,5-disulphonate (tiron) as substrate. The iron(III) reacted with tiron to form a complex which absorbs strongly at 667 nm, where it was measured spectrophotometrically. The system has a linear range of 1 to 50 mg L–1 with a detection limit of 0.67 mg L–1. It is suitable for the determination of total iron in multivitamin tablets and iron-rich ground waters, with a relative standard deviation of better than 1.1%. The results obtained compared favourably with the certified values and a standard ICP-AES method.
Published Version
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have