Abstract
The pulsed flame photometric detector (PFPD) has the advantage of being able to measure the concentration of individual sulfur compounds and total sulfur content in a petrochemical sample in a single gas chromatography run. Because it is an equimolar response detector, the PFPD's sulfur response is independent of a compound's molecular structure, and this feature allows quantitation of the total or speciated sulfur content in complex samples using a single calibrant. This paper is a survey describing a variety of applications using the PFPD for sulfur quantitation in petrochemical matrices. Several different approaches to quantitation are described, and simple techniques for circumventing the quenching of the sulfur signal by coeluting hydrocarbon peaks are discussed. Examples from a range of real-world samples are presented.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.