Abstract

A high-pressure liquid chromatographic method for determining the concentrations of ticarcillin in serum was developed and compared, with 93 patient sera, to a standard agar well diffusion bioassay. For analysis, serum plus temocillin, the internal standard, were extracted with chloroform-n-amyl alcohol and back extracted into phosphate buffer. A reverse-phase C18 column and an ammonium acetate-methanol mobile phase were used with detection at 242 nm. Reproducibility studies yielded coefficients of variation ranging from 2.4 to 4.7% for low, mid, and high controls. Although cefoxitin, cephalothin, and cefuroxime exhibited retention similar to that of ticarcillin, a wide range of commonly administered antibiotics and other drugs did not interfere. The high-pressure liquid chromatographic assay is an accurate, reproducible method for determining the concentration of ticarcillin in serum during multiple antibiotic therapy or when rapid results are required.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.