Abstract
Endocrine-disrupting compounds are routinely detected in environmental water. In this study, a simple and rapid sample preparation method based on ultrasound-assisted salt-induced liquid-liquid microextraction (UA-SI-LLME) followed by high performance liquid chromatography – diode array detection (HPLC-DAD) was developed for the determination of triclosan, triclocarban and methyl-triclosan in water. Parameters that affect the extraction efficiency – such as type and mass of salt, type and volume of extraction solvent, extraction time, sample pH and sample volume – were studied. Under the optimized conditions, the method demonstrated linearity from 0.18 to 1000 µg L−1 with coefficients of determination R2 ≥ 0.9899, limits of detection from 0.05 to 0.2 µg L−1, and good recoveries (> 70%) with relative standard deviations < 10%. The method was applied for the determination of triclosan, triclocarban, and methyl-triclosan in swimming pool, lake, and wastewater. The greenness of the method was quantified using the Analytical Eco-Scale Index.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.