Abstract

Chromatographic separation of methylxanthine metabolites was achieved using a Hypersil octadecylsilane column with a simple concave gradient elution programme of 0–12.75% acetonitrile in 1% tetrahydrofuran, pH 4.8, and the eluted components were detected by monitoring their absorption at 280 nm. An extraction procedure involving the formation of an ion-pair complex was developed which gave significant improvements over previously described methods including a shorter chromatographic run of 20 min. A thorough comparison of this procedure with a more convenient alternative involving direct injection of diluted urine specimens showed that the latter analysis was adequate for the quantitation of the major urinary metabolites of caffeine and theophylline.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.