Abstract

A reactor neutron activation analysis procedure for the determination of the silver content of silver coins is described. The samples are irradiated for 1 s, cooled for 85 s and measured for 60 s with a Ge-detector. The analysis is based on the measurement of110Ag and108Ag. Aluminium is used for flux monitoring and pulse pile-up correction. A calibration curve is prepared by irradiating and measureing a series of discs with known silver contents. An average precision of ±2.1% is obtained. The analysis of coins with known silver contents shows good agreement with the given values. The analysis time is 2.5 minutes per sample.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.