Abstract

LiF based thermoluminescent materials are the most used passive detectors in the field of radiation protection. For personal and area dosimetry in mixed neutron and photon fields, /sup 6/LiF and /sup 7/LiF TLD pairs are employed for the simultaneous determination of neutron and photon doses. Therefore, an accurate knowledge of their sensitivity to photons and thermal neutrons is essential. Aimed at providing reliable sensitivity data for new LiF based TL materials, an experimental study was performed in the framework of collaboration between INFN Frascati National Laboratories and ENEA, Italy. The /sup 6/LiF(Mg, Cu, P) material from TLD Poland, was specially manufactured in six versions with different sensitive thickness (t=40,60,80,100,150 and 230 mg/spl middot/cm/sup -2/), which implies different neutron and photon sensitivity. Together with the /sup 7/LiF(Mg, Cu, P) material, namely MCP-7, all seven type of TLDs were irradiated in reference beams of photons and thermal neutrons, at the calibration facilities of INFN and ENEA. The study provides a set of sensitivity data and discusses the performance of the investigated materials for the dosimetry of mixed neutron and photon fields. All uncertainties or error bars reported in the paper refer to 95% confidence level (2 sigma).

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call