Abstract

Since 1998 neutron dosimetry based on PADC (poly allyl diglycol carbonate) is done with a so-called original Paul Scherrer Institute (PSI) design at PSI. The original design (i.e. holder) was later changed. Both designs are optimised for use in workplaces around high-energy accelerators, where the neutron energy spectra are dominated by fast neutrons ranging up to some 100 MeV. In addition to the change of the dosemeter design a new evaluation method based on a microscope scanning technique has been introduced and the etching conditions have been optimised. In the present work, the responses obtained with the original and the new dosemeter designs are compared for fields of radionuclide sources and monoenergetic reference fields using the new evaluation method. The response curves in terms of the personal dose equivalent for normally incident neutrons were built as functions of the incident neutron energy.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.