Abstract

Flame-spectrophotometric and atomic-absorption methods of assaying trace amounts of thallium in biological material have been investigated. With the Perkin-Elmer, Model 303, spectrophotometer atomic absorption is the more sensitive and yields detection limits comparable with those reported in the literature for the most sensitive types of flame-emission equipment. A simple and rapid method of micro sampling by using a tantalum boat is described. It enables nanogram amounts of thallium (down to 3 ng) in 50 to 100 µl of blood and urine to be determined with an accuracy and precision of 3 to 5 per cent. This represents an increase in sensitivity of at least twenty-five times over conventional atomic-absorption spectrophotometry. The results obtained with the tantalum boat are susceptible to inter-element interferences, and calibration by using the additions method is essential for accurate quantitative results.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.