Abstract

A method is described for the determination of tellurium at nanogram levels in rocks and in other complex materials by the use of flameless atomic-absorption spectroscopy. A very selective organic extraction procedure is applied to avoid matrix interference effects during extraction of Te and the atomization stage in the graphite furnace. Prior separation of iron and other interfering elements is achieved by a combined cupferron—ethyl acetate extraction. Tellerium is extracted from 6 M hydrochloric acid with MIBK and stripped into aqueous medium. Pipetting of the aqueous extract into the graphite furnace gives fairly good instrumental reproducibility (2–3% error). Detection limits of about 10 ppM Te for a 0.5-g sample have been achieved with the medium-performance apparatus used. Results for Te in some geochemical reference materials are reported. Indications are given for the determination of Sb and Mo in the same solutions.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.