Abstract
A method for the determination of 1–500 p.p.m. of acid-soluble aluminium and 2–500 p.p.m. of acid-insoluble aluminium in low-alloy and stainless steels by flameless atomic absorption with a HGA 74 graphite furnace is described. A typical value of the relative standard deviation for acid-soluble aluminium at concentrations larger than 10 p.p.m. was 5 %. The steel sample was dissolved in hydrochloric and nitric acid and filtered. Ammonium sulphate was added to the filtered sample in order to overcome interferences caused by hydrochloric acid. No concentration steps were used. Acid-soluble aluminium was determined within 20 min. The influence of iron, chromium, nickel, molybdenum, hydrochloric acid, nitric acid and ammonium sulphate respectively was investigated.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.