Abstract
Abstract Three analytical wavelengths (180.731,182.037, and 182.625 nm) were tested for determination of sulfur in 2 fertilizers by inductively coupled plasmaatomic emission spectrometry (ICP-AES). Variable S results at 182.037 nm were caused by a downward sloping background shift (Fe line at 181.851 nm) during measurement of the sample solutions. At 182.625 nm an emission peak doublet, 182.619/182.635 nm, was observed from which the instrument selected the measurement peak. A shift from calibrated peak 182.619 nm to noncalibrated peak 182.635 nm occurred when the S/B ratios were ≤22-31 in the fertilizer samples examined. For the 3 wavelengths of S investigated, the pattern of spectral and interelement interference changed at the same wavelength according to the composition of the fertilizer. K, P, and Mg typically caused interelement effects and Ca and B caused spectral effects during S determination. In the interference studies, 180.731 nm proved the best wavelength for the ICP-AES determination of S in fertilizers. Interlaboratory studies showed the accuracy of the method to be ± 6.2% for both water- and acid-soluble S, independent of the measurement wavelength, instrument, and operator.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.