Abstract

Determination of suitability test on digital mammography have been conducted from 2016-2018 at a hospital in Makassar City using voltage accuracy parameter. This measurement was used detector, and X-ray multimeter. Statistically, voltage accuracy parameter has correlation value, r = 0.866 (perfect correlation). While the p value obtained in voltage accuracy parameter is p < 0.05. There are significant differences in value from 2016-2018. The result of the suitability test for voltage accuracy parameter is still within the tolerance limit as recommended by the IAEA.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.