Abstract

Atomic absorption spectrometry with an induction furnace is used for the determination of silver (0.03–12 μg g -1) and thallium (0.004–12 μg g -1) in 1–40-mg samples of nickelbase alloys dropped into the furnace. Calibration graphs of peak absorbance versus mass of element were constructed by use of standardised alloys. The accuracy, precision and limits of detection of the method are described for numerous nickel-base alloys. With alloys containing more than 0.2 μg Ag g -1 and 0.4 μg Tl g -1, relative standard deviations are usually <15% and <12%, respectively. The limits of detection for silver and thallium are 0.014 μg g -1 and 0.004 μg g -1, respectively.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.