Abstract
Atomic-absorption spectrometry with a resistively heated furnace has been used for the determination of 5–52 µg g–1 of arsenic, 0.1–5 µg g–1 of antimony, 0.07–20 µg g–1 of selenium and 0.07–2.1 µg g–1 of tellurium in milligram samples of nickel-base alloys added directly to the furnace. Calibration graphs of peak area versus the mass of trace element have been constructed by using standard nickel-base alloys. Information is presented on the accuracy and precision of the method for 23–34 alloys depending on the element under investigation. The limits of detection are arsenic 0.1, antimony 0.1, selenium 0.06 and tellurium 0.003 µg g–1.Results are also presented for the determination of 1–8 µg g–1 of tellurium in nickel-base alloys using atomic-absorption spectrometry with the introduction of solid samples to an induction furnace and for the determination of arsenic in the nickel-base alloys using atomic-absorption spectrometry after hydride generation and with electrothermal atomisation of solutions of the alloys.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.