Abstract

We aimed to measure propofol concentrations in exhaled air with an ion mobility spectrometer coupled to a multicapillary column for pre-separation (MCC-IMS). In addition, we aimed to compare the values of these measurements with serum propofol concentrations, as determined by gas chromatography-mass spectrometry (GC-MS). Thirteen patients, ASA I or II, undergoing elective ENT surgery were studied. Anaesthesia was induced with propofol 2.1 (0.7) mg kg(-1), rocuronium 0.5 (0.1) mg kg(-1), and remifentanil 0.5 microg kg(-1) min(-1). After tracheal intubation, anaesthesia was maintained with a continuous infusion of propofol 3.9 (1.8) mg kg(-1) h(-1) and remifentanil 0.5 microg kg(-1) min(-1). Simultaneously, a venous blood sample was obtained. Propofol concentrations in serum were determined by GC-MS and compared with the height of the respective propofol signals achieved by MCC-IMS. Twenty-four pairs of samples were obtained. The comparison of propofol concentrations in exhaled air and serum presented a bias of -10.5% and a precision of +/- 12.3%. With these values, the 95% limits of agreement were 14.1% and -35.1%. MCC-IMS may be a suitable method to determine propofol concentrations in exhaled air, and may be used to predict propofol concentrations in serum.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.