Abstract

Methyl iodide has been highlighted as a potential genotoxic impurity (PGI). A sensitive HS-GC method with ECD was developed and validated for the determination of methyl iodide impurity in Tedizolid phosphate, an active pharmaceutical ingredient. HS-GC method on DB-624,column 60m X 0.53mm, film thickness 3μm, with electron capture detector (ECD) was used. The proposed method was specific, linear, accurate, rugged and precise. The calibration curves showed good linearity over the concentration range of 0.78 μg/g to 11.7 μg/g and the correlation coefficient was 0.997. Method had very low limit of detection (LOD) and limits of quantification (LOQ) are 0.23 μg/g and 0.78 μg/g, respectively, with respect to test concentration. Accuracy was observed within the range 96.0% to 103.2%. This method is a further extended good quality control tool for low level quantitation of methyl iodide impurity in another API, alogliptin benzoate.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.