Abstract

Abstract A high-performance liquid chromatographic method for determining reserpine in plasma has been developed. The procedure involves extraction of reserpine from buffered plasma into benzene, oxidation of reserpine to a fluorophor by treatment with vanadium pentoxide in phosphoric acid, and chromatographic separation of the reserpine fluorophor on an octadecylsilane column by ion-pairing with heptanesulfonate ions. Fluorescence monitoring of the column effluent provides high sensitivity of detection and increases the specificity of the procedure. A detection limit of approximately 100 pg of reserpine per ml of plasma was obtained following analysis of 2 ml samples. Analysis of a number of samples demonstrated the applicability of this method in confirming the presence of reserpine in equine plasma specimens collected at various horse shows and in evaluating the pharmacokinetic behavior of reserpine following intramuscular administration to horses.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.