Abstract
A heat clearance method is described for measuring regional cerebral cortical blood flow, involving the injection of a small amount of cool saline into the cerebral circulation. Blood flow is estimated from the slope of the subsequent thermal clearance curve, temperature being measured by a small thermistor probe situated under the dura in contact with cerebral cortical tissue. The method is compared with the well established technique involving 85Kr an a correlation coefficient of 0.89 is obtained. The heat clearance technique, however, has the added advantages that the cost of tracer and monitoring equipment is small and that probes may be chronically implanted.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.