Abstract
Determination of plutonium by secondary coulometry involving the controlled potential technique for the generation of an iron(II) mediator, is reported in Part I.* In this paper, the same determination is reported using constant current coulometry for the generation of the mediator, and zero current potentiometry for end-point detection. The factors affecting the current efficiency, viz. current density and supporting electrolyte composition have been checked in order to define the appropriate conditions for obtaining 100% current efficiency. The original method of Carson et al. suffers from the disadvantage that it involves complicated sample treatment. Introduction of perchloric acid treatment as a method for the oxidation of plutonium and the pretitration of the supporting electrolyte to the end-point potential prior to sample addition, have considerably helped improve the precision and accuracy of the method. Exhaustive analytical data are reported covering plutonium quantities ranging from 25 micrograms to 5 milligrams, which establishes the scope of the method.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.