Abstract
The photon backscatter factor for several irradiation phantoms was determined by calculational and experimental methods and compared with a phantom constructed of International Commission on Radiation Units and Measurements (ICRU) four-element tissue. Calculations of the photon backscatter factor over the energy range from 10 to 2000 keV were performed using the MCNP 4A code for each of the phantoms. Measurements of the backscatter factor were carried out using a thin-walled ionisation chamber that was exposed to X-ray beams with narrow energy spectra. The calculations and measurements agreed to within uncertainties and demonstrated that the backscatter factors over this energy range for a water-filled phantom recommended by the International Organization for Standardization (ISO) and a phantom constructed of tissue-equivalent (RS-I) plastic are nearly the same as that of the ICRU tissue reference phantom. However, the backscatter from a polymethylmethacrylate (PMMA) phantom was up to about 8% higher. In addition, it was found that a composite phantom of polytetrafluoroethylene (PTFE) and PMMA also produced a backscatter factor quite similar to that of the ICRU tissue reference phantom.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.