Abstract

A method for sampling and analysis of phenolic compounds in workplace atmospheres using Amberlite XAD‐2 sorbent tubes has been laboratory and field tested. After extraction of the adsorbed phenols with methylene chloride, the concentrated extract was analyzed by capillary column gas chromatography coupled with a flame ionization detector or a mass spectrometer. For a 3L air sample, the method has a detection limit for phenol, cresols, alkyl‐substituted phenols and xylenols of 0.3 mg/m3 using gas chromatography with flame ionization detection (GC/FID) analysis and 0.07 mg/m3 using gas chromatography mass spectroscopy/selected ion monitoring GC‐MS/SIM. Evaluation tests conducted in the laboratory indicated recoveries >85% for a wide range of phenols, cresols, xylenols and other alkyl‐substituted phenols and resorcinol. Sampling conditions were established to ensure that no breakthrough occurred for a predetermined sample volume of three (3) liters even in humid atmospheres. Stability tests indicated that fo...

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.