Abstract

In this work, a method to determine Pb(II) ions in model water is presented; the method is based on the fluorescence emission of a silver nanoclusters (AgNCs) colloidal solution, which is sensitive to lead ions. The presence of Pb(II) ions causes a photoemission enhancement of the AgNCs solution dependent on the pollutant concentration. The functional dependence is logarithmic in the range from 2.5 to 40 µM, and through the linearization of the calibration points, a linear function is determined and exploited for the extrapolation of the test Pb(II) concentrations with a precision estimated by relative standard deviation (RSD) ranging from 21% to 10% from the highest to the lowest Pb(II) quantity, respectively. Finally, inductively coupled plasma–optical emission spectroscopy (ICP-OES) successfully validated the described method. The accuracy of the method is also studied for intentionally polluted mineral waters, revealing the same trend of the model water: the lower the concentration, the higher the precision of the method.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.