Abstract
A method for the determination of 24 organochlorine pesticides by gas chromatography-mass spectrometry (GC-MS) with negative-ion chemical-ionization (NICI) using programmable temperature vaporizer (PTV)-based large volume injection (LVI) is described. The ion source temperature was determined to be 150 °C for the optimized NICI-selected ion monitoring (SIM) conditions. PTV inlet parameters were also optimized. The sensitivities of the pesticides by splitless-GC-NICI-MS were approximately 7.8–360 times higher than those of the pesticides by splitless-GC-EI-MS. The sensitivities of the pesticides by LVI-GC-NICI-MS were over 80 times higher than those of the pesticides by spiltless-GC-NICI-MS. This method was applied to the determination of the pesticides in river water using micro liquid–liquid extraction as sample preparation. The recoveries of the pesticides from a river water sample spiked with standards at 2 ng l−1 and 20 ng l−1were 75–111% (RSD, 2.9–15%) and 92–105% (RSD, 0.5–5.6%), respectively. The methodical detection limits ranged from 0.004 to 2.2 ng l−1
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.