Abstract

A simple and rapid fluorescence polarization immunoassay (FPIA) based on polyclonal antibodies is reported for the first time for the determination of nicotinic acid. First, six nicotinic acid analogues were coupled with fluorescein isothiocyanate isomer I (FITC) to obtain tracers. The influences of tracer structure on the performance of FPIA as well as the structure-activity relationships were explored. Subsequently, the tracer concentration and antibody dilution ratio on the performance of FPIA were investigated. For the optimal conditions, the detection range of FPIA was from 1.1 to 145.6 μg mL−1 and its IC20 value as the limit of detection (LOD) was 1.1 μg mL−1. The cross-reactivity of antibodies with common vitamins was evaluated. The intra-assay recovery was from 92.0% to 101.2% in food and tablets, with inter-assay recoveries between 92.0% and 103.8%. The developed FPIA provided good sensitivity and reproducibility and therefore provided efficient and economic screening for nicotinic acid in food and tablets.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.