Abstract
A sensitive high-performance liquid chromatographic method for a routine assay of nadolol in serum is described. Serum samples spiked with atenolol (internal standard) were extracted with diethyl ether. After centrifugation, the organic layer was evaporated to dryness. The residue was redissolved in the mobile phase and injected onto an octadecyl silica column (150 mm x 4.6 mm I.D.). The mobile phase was 0.05 M ammonium acetate (pH 4.5)-acetonitrile (85:15, v/v). Fluorometric detection (excitation 230 nm, emission 300 nm) was used. The minimum detectable level of nadolol in serum was 1 ng/ml.