Abstract

In order to determine metal ions in aqueous solution, a millisecond pulsed solution cathode glow discharge (SCGD) source coupled with a portable spectrometer was developed for atomic emission spectrometry (AES). The millisecond pulsed SCGD source was operated with two high-voltage DC power supplies and a high-voltage pulse switch, specifically boosting the analyte excitation and sustain the glow discharge. The method to excite the plasma discharge contrasts with traditional approaches that employ a single power source. The influence of operating parameters, including pulsed voltage, pulsed frequency, duty ratio, and sample flow rate, were optimized to construct a system having good stability and improved excitation of metal ions. The optimum detection limits for Na, K, Li, Mg, Ag, Cs, Rb, and Sr were 0.17, 0.32, 0.38, 0.50, 2.65, 1.30, 0.22, and 74.02 µg·L−1, respectively. Furthermore, mountain spring, Yangtze River, and reservoir water were spiked with Ag, Rb, and Cs; the recoveries ranged from 105% to 106%, 106% to 107%, and 107% to 108% indicating that the developed system has favorable analytical performance.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call