Abstract
The paper describes the determination of ultratrace concentrations of mercury in biological materials. Samples were digested in nitric acid under pressure in a polytetrafluoroethylene bomb. Mercury was determined in the digest solution by cold vapour atomic absorption spectrometry using stannous chloride reduction and two-stage gold amalgamation. The detection limit is 0.12 ng. Analysis of several certified reference materials yielded results in excellent agreement with the certified values. The results for blood serum of several subjects ranged from 0.2 to 0.8μg 1−1.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.